AD-TV 200 GS

Description

The isolation amplifier AD-TV 200 GS is used for galvanic isolation, amplification and conversion or adaptation of analog measurement signals. Two front-side spindle trimmers can be used to additional adjust the offset (zero) and full scale (span). The integrated wide-range power supply allows operation in a wide supply voltage range with low self-heating.

AD-TV 200 GS

Application

Galvanic isolation, adapt, amplify and transform analog measurement signals.

Business data

Order number
AD-TV 200 GS
Identifiers
EClass 27210120
Customs tariff number / country of origin 8543 70 90 / Deutschland

Information

Downloads
Data sheet pdf en-tv200gs.pdf (156 kB)
Tender text zip tv200gs.zip (13 kB)
Safety instructions pdf ad-safety-instructions.pdf (43 kB)

Technical specifications

Input
Measuring range current 0 ...20 mA; 4 ... 20 mA1)
Input resistance 50 Ohm
Measuring range voltage 0 ... 10 V, 2 ... 10 V1)
Input resistance 100 kOhm
Output
Output range current 0 ... 20 mA; 4 ... 20 mA1)
Max. burden 500 Ohm
Output range voltage 0 ... 10 V; 2 ... 10 V1)
Min. burden 1 kOhm
Residual ripple < 50 mVss
1)adjustable; others on request
Housing ST
Dimensions (WxHxD) 23 x 78 x 103 mm
Type of protection IP 20
Connection method Screw clamp
Terminals, wire cross section 2,5 mm² flex wire / 4 mm² one wire
Bolting torque terminals 0,5 Nm
Weight ~ 120 g
Manner of fastening 35 mm DIN rail 35mm
Transmission behaviour
Basic accuracy < 0,3 %
Temperature influence < 100 ppm/K
Supply
Supply voltage 20...250 V DC / 50...250 V AC
Nominal voltage 24 V DC / 230 V AC
Power consumption Max 2 W / 3 VA
Environmental conditions
Ambient temperature -10 ... 50 °C
Storage and transport -10 ... 70 °C (no condensation)
EMC
Product family standard EN 61326-1 2)
Emitted interference EN 55011, CISPR11 Cl. B, Gr. 1
2) During electromagnetic disturbance minor changes in output signal are possible.
Electrical safety requirements
Product family standard EN 61010-1
Overvoltage category II
Pollution degree 2
Galvanic isolation, test voltages
Input/output 3,75 kV RMS (1 min.)
Signal/auxiliary voltage 3 kV RMS (1 min.)
Protective systems
Input/output Electrical surge protection
Power supply Electrical surge and reverse current protection

Block and wiring diagram

AD-TV 200 GS

Dimensions

AD-TV 200 GS

CAD / CAE Daten

CAD model is not yet available for this product.

The CAD product overview can be found at: adamczewski.partcommunity.com