AD-MV 50 GX, AD-MV 54 GX

Description

The measuring value amplifier AD-MV 50 GX (for 2/3 lead technology) or AD-MV 54 GX (for 4-lead technology only) converts the measuring signal issued from a resistance generator Pt 100 (according to DIN IEC 751) to an impressed output signal (i.e. 0-20 mA or similar). The output characteristic curve is issued temperature linear. When using the 3-lead switching lead, resistance up to 100 Ohms are compensated. In 2-lead switching a subsequent zero alignment and full alignment via trim potentiometers at the front is possible. The analogue output signal is galvanically separated from the supply voltage. A highly efficient, integrated electronic wide-range power pack (ALLPOWER) enables operation with 18-30V DC. The power supply is independet of polarity.A high packing density is achieved due to its narrow build.
Temperature range and output must be defined when ordering equipment.

AD-MV 50 GX, AD-MV 54 GX

Application

Temperature measuring with Pt 100 temperature transmitter for continuous transformation of temperature variables to analogue signals.

Specific characteristics

  • Pt 100 signal conversion into standard signals
  • wide range power supply
  • conversion of 2, 3-wire technology (MV50GX)
  • conversion of 4-wire technology (MV54GX)
  • narrow design
  • operating display with green LED

Business data

Order number
AD-MV 50 GX For 2 - and 3-wire connection
AD-MV 54 GX For 4-wire connection
Accessory
DIN-rail connector AD-GX Connector
Identifiers
EClass 27210129
Customs tariff number / country of origin 8543 70 90 / Deutschland

Information

Technical specifications

Pt100 input
Max. measuring range -200 ... 850 °C1)
Connection method 2, 3 or 4-wire
Output current
Output range 0 ... 20 mA, 4 ... 20 mA1)
Max. burden 400 Ohm
Residual ripple 20 µAss
Output voltage
Output range 0 ... 10 V1)
Min. burden 10 kOhm
Residual ripple 10 mVss
Supply
Voltage range DC 18 ... 30 V DC
Nominal voltage DC 24 V DC
Power consumption DC 650 mW
Transmission behaviour
Accuracy < 0,2 %
Linearity error < 0,2 %
Temperature influence 100 ppm/K
Response time ~ 50 ms
Housing
Dimensions (WxHxD) 6,2 x 92 x 101 mm³
Type of protection IP 20
Connection method Screw clamp
Terminals, wire cross section 2,5 mm² flex wire / 4 mm² one wire
Bolting torque terminals 0,5 Nm
Weight ~ 70 g
Manner of fastening 35 mm DIN rail 35mm
Environmental conditions
Ambient temperature -10 ... 50 °C
Storage and transport -10 ... +70 °C (avoid condensation)
EMC
Product family standard 2) EN 61326-1
Emission 3) EN 55011, CISPR11 Cl. A, Gr. 1
Electrical safety requirements
Product family standard EN 61010-1
Overvoltage category II
Pollution degree 2
Galvanic isolation, test voltages
Signal / supply unit 1,5 kV (1 min.)
Protection circuits
Input Electrical surge protection
Output Electrical surge protection
Power supply Electrical surge and reverse current protection
1) Specify signal range in plain text when ordering.
2) During electromagnetic disturbance minor changes in output signal are possible.
3) Warning:
This device is not intended to be used in residential areas and can not ensure adequate protection of radio reception in such environments.

Block and wiring diagram

AD-MV 50 GX, AD-MV 54 GX

Dimensions

AD-MV 50 GX, AD-MV 54 GX

CAD / CAE Daten

CAD model is not yet available for this product.

The CAD product overview can be found at: adamczewski.partcommunity.com

Display and operating elements

Designation Meaning
On LED green, Power supply
N Trimmer 0 %-Value
V Trimmer 100 %-Value
Cookie settings